XRD Dental Equipment ← Back to Tech Library
Installation GuideImagen Sensor
XRD Imagen Intraoral Sensor · Direct Integration

Using your own imaging software? 6 steps and you're shooting.

This guide takes your Imagen sensor from the box to acquiring images directly in your practice imaging software. Two items, six steps, about 10 minutes — no IT experience required.

About 10 minutes
Windows 10 or 11 PC · free USB port
No IT knowledge needed

The one rule that matters: plug things in at the right time.

Only two USB devices this time. Each one connects at a specific step — this panel lights up as you complete them.

STEP 2
Installer flash drive
Installer flash drive
TWAIN driver + calibration
🔒 Waits its turn
STEP 4
Imagen sensor
Imagen sensor
Connects after the driver
🔒 Waits its turn
0 of 6 steps done
Before you start

What's in the box

For a TWAIN installation you only need two items from your Imagen box, plus a Windows 10 or 11 computer with a free USB-A port and your practice imaging software already installed.

Imagen sensor with cable Imagen sensorWith its white protective cover — keep the cover on whenever the sensor is not in use.
Installer flash drive Installer flash driveContains the TWAIN driver and your sensor's own calibration files, matched to its serial number.

Golden rule: the driver goes in before the sensor. Don't connect the sensor until Step 4 — plugging it in early is the #1 cause of installation problems. Already plugged it in? No harm done: unplug it and start from Step 1.

Installation

The 6 steps

Tap each step when you finish it — the bar at the top tracks your progress.

Close all open programs, including your imaging software. If your antivirus is very aggressive, pause it during the installation (you can turn it back on right after).

Make sure you're logged into Windows as an administrator account.

Nothing from the Imagen box should be connected to the PC at this point.

Insert the installer flash drive into a USB port. Do not connect the sensor yet.

Open the drive in File Explorer and locate the TWAIN folder.

Quick check: the serial number printed on the installer drive should match the one on your sensor's USB connector — the calibration files on this drive belong to that specific sensor. Keep the drive in a safe place after the installation.
Installer flash drive

Inside the TWAIN folder, right-click the setup file and choose "Run as administrator".

Follow the wizard — accept and keep clicking through until it finishes. The driver and your sensor's calibration files are installed automatically.

Good to know for later: if this driver ever gets corrupted, running the same setup again offers a "Repair" option that fixes it in place.

Plug the Imagen sensor into a free USB-A port on the computer itself (avoid USB hubs). Always hold it by the connector — never pull or twist the cable.

Windows will recognize it in a few seconds.

Imagen sensor

Open your imaging software and go to its acquisition or device settings. Choose TWAIN as the image source, then pick the Imagen sensor from the device list.

Every program words this a little differently — it's usually called something like "Select TWAIN source" or "Acquire from TWAIN". If you can't find it, email us a screenshot of your software's settings and we'll point to the exact spot.

Slide a disposable hygiene sheath over the sensor (a new one for every patient).

Start an acquisition from your imaging software, position the sensor with the flat side facing the X-ray source, and press the exposure button on your X-ray unit. The recommended range is 60–70 kV — if you shoot with the XRD 70, its 70 kV output is a perfect match.

Your image appears in your software within seconds. That's it — you're fully installed.

Use a sensor positioner if you have one — it keeps the sensor parallel to the tooth and gives you a clean first image.
Veterinary teams: dental radiographs on animals are normally taken under sedation or anesthesia. Keep the sensor parallel to the tooth root, and never leave it in the mouth unattended — a single bite can permanently damage the chip inside.
🎉 All done! Your Imagen sensor is installed and ready to acquire from your imaging software. Keep the installer flash drive and this guide somewhere safe.
If something looks off

Quick fixes

Nine times out of ten it's one of these. Each fix takes under five minutes.

Your software can't see the sensor

First, the obvious: check the sensor is firmly plugged in — and try another USB port on the computer itself (avoid USB hubs). Then confirm you selected TWAIN as the source and the Imagen sensor as the device (Step 5).

If it still doesn't appear: open Windows Device Manager and check whether the sensor is listed. If it isn't recognized, run all pending Windows Updates (including optional ones), temporarily disable antivirus and firewall, restart the PC, and run the TWAIN setup again from the installer flash drive. Then test again.

The image freezes, cuts off, or looks scrambled

This is almost always Windows putting the USB controller to sleep. Open Device Manager → System devices → Intel(R) Management Engine Interface, right-click Properties → Power Management, and uncheck "Allow the computer to turn off this device to save power". Click OK and try again.

Driver errors or missing calibration messages

Run the TWAIN setup again from the installer flash drive and choose "Repair" when the option appears — it restores the driver and the calibration files in place. If the message persists after the repair, email us before trying anything else.

Installing more than one sensor on the same computer

Each Imagen sensor has its own calibration files. If you use two or more sensors on one PC, run the installer once for each sensor's serial number — the calibration for each unit is installed automatically.

Still stuck? Send us the details

Email us with: your sensor's serial number (printed on the flat side of the USB connector), your Windows version, the name of your imaging software, what the error says, whether this is a first installation, and if possible a photo or screenshot. With that, we can usually solve it in one reply.

Support: info@xrddental.com

Make it last for years

Caring for your Imagen sensor

New hygiene sheath for every patientDisposable covers sized for your sensor. Never reuse them.
Disinfect with 70% isopropyl alcoholWipe the sensor and cable with a moistened cloth before each use. No other liquids.
Hold the connector, not the cableWhen plugging and unplugging, always grip the USB connector itself.
Keep the protective cover onBetween uses, store the sensor in its cover, away from static and drops.
Never autoclave, UV or dry heatThe sensor cannot be heat-sterilized — it would be destroyed instantly.
Don't bend, twist or yank the cableCable damage is the most common cause of sensor failure — and it's avoidable.
No bites, drops or impactsThe CMOS chip inside is precise — and fragile. Human or animal, no biting down on the sensor.
Never open the sensorThere are no serviceable parts inside. If anything seems wrong, write to us first.

We've got your back.

More videos, guides and product material are in our Tech Library — and if anything in this setup gives you trouble, one email is all it takes.

Email support Tech Library