XRD Dental Equipment← Back to Tech Library
Installation GuideSensor + ArchiMED
XRD Imagen Intraoral Sensor · ArchiMED Suite 5

From the box to your first X-ray, in 8 easy steps.

Your Imagen sensor was designed to be truly plug-and-play. Follow these steps in order — the whole setup takes about 15 minutes, and no IT experience is required. The same steps apply to both the human and the veterinary edition of ArchiMED.

This guide covers the complete Sensor + ArchiMED installation. Using your own imaging software instead? Follow the sensor installation guide (direct integration).
About 15 minutes
Windows 10 or 11 PC · free USB port
No IT knowledge needed

The one rule that matters: plug things in at the right time.

Your kit has three USB devices. Each one connects at a specific step — this panel lights up as you complete them.

STEP 2
Installer flash drive
Installer flash drive
Software + drivers + calibration
🔒 Waits its turn
STEP 5
License dongle
License dongle
Stays connected — always
🔒 Waits its turn
STEP 7
Imagen sensor
Imagen sensor
Connects last
🔒 Waits its turn
0 of 8 steps done
Before you start

What's in the box

Open your Imagen box and check that all three items are there. You'll also need a Windows 10 or 11 computer with a free USB-A port.

Imagen sensor with cable Imagen sensorWith its white protective cover — keep the cover on whenever the sensor is not in use.
Installer flash drive Installer flash driveContains the ArchiMED software, drivers, manuals and your sensor's own calibration files.
License dongle License dongleYour software license lives on this key. It is unique and personal — store it like a valuable.

Golden rule: start with nothing plugged in. Don't connect the sensor or the license dongle yet — each device has its moment, and this guide tells you exactly when. Plugging them in early is the #1 cause of installation problems. Already plugged something in? No harm done: unplug it and start from Step 1.

Installation

The 8 steps

Tap each step when you finish it — the bar at the top tracks your progress.

Close all open programs. If your antivirus is very aggressive, pause it during the installation (you can turn it back on right after).

Make sure you're logged into Windows as an administrator account.

Nothing from the Imagen box should be connected to the PC at this point.

Insert the installer flash drive into a USB port. Do not connect the sensor or the license dongle yet.

Open the drive in File Explorer, right-click Setup and choose "Run as administrator".

This drive carries your sensor's personal calibration files. Quick check: the serial number printed on the installer drive should match the one on your sensor's USB connector. Keep the drive in a safe place after the installation.
Installer flash drive

In the installer window, click "Program Installation", then Next.

Accept the license terms and click Next again. The software installs itself — just wait and click Finish at the end.

The wizard makes every choice for you — there is nothing to configure here. Some installer screens may show the name of our European software development lab; that's normal and expected.

Right after the program, the installer offers to set up the drivers for the USB license key. Accept and click Next.

Windows will ask to add a firewall exception — click "Yes". This is expected and safe; without it, the license key cannot be recognized.

Wait for it to complete and click Finish.

Insert the license dongle into a free USB port. It holds your software license and must stay connected whenever you use the software.

Best practice: use a rear USB port on a desktop PC, where the dongle is protected and out of the way.

Treat this key like a valuable. The license is personal — if the dongle is lost, a new license must be purchased.
License dongle

Open ArchiMED from the desktop icon. On the very first launch, the software asks you to create the image database: select the disk where your X-ray images will be stored (drive C: is fine) and press Add. Don't skip this — without a database, no image can be saved.

Sign in as Admin with the default password 12345678. The software will immediately ask you to create your own secure password — this protects your patient data.

Then fill in your practice information when prompted. Enter it carefully: this data is stamped into the clinical record of every image you acquire.

Plug the Imagen sensor into a free USB-A port on the computer itself (avoid USB hubs). Always hold it by the connector — never pull or twist the cable.

Windows will recognize it in a few seconds. In ArchiMED, choose the Intraoral category, enter the Setup menu, select your sensor model from the device list, and save.

Go back with Previous — a pop-up window appears, and if the installation is correct the logo turns green. Green logo = your sensor is ready to shoot.

Your sensor's calibration files were installed automatically back in Step 3 — they match its serial number, so there's nothing to calibrate manually.
Imagen sensor

Slide a disposable hygiene sheath over the sensor (a new one for every patient).

Create a test patient in ArchiMED (the orange fields are mandatory), then double-click the patient's name to open the acquisition screen. Position the sensor with the flat side facing the X-ray source, and press the exposure button on your X-ray unit. The recommended range is 60–70 kV — if you shoot with the XRD 70, its 70 kV output is a perfect match.

Your image appears on screen within seconds. That's it — you're fully installed.

Use a sensor positioner if you have one — it keeps the sensor parallel to the tooth and gives you a clean first image.
Veterinary teams: dental radiographs on animals are normally taken under sedation or anesthesia. Keep the sensor parallel to the tooth root, and never leave it in the mouth unattended — a single bite can permanently damage the chip inside.
🎉 All done! Your Imagen sensor is installed and ready for clinical use. Keep the installer flash drive and this guide somewhere safe — and remember: the license dongle stays plugged in.
If something looks off

Quick fixes

Nine times out of ten it's one of these. Each fix takes under five minutes.

The software says the sensor is not connected

First, the obvious: check the sensor is firmly plugged in — and try another USB port on the computer itself (avoid USB hubs).

If it still doesn't appear: open Windows Device Manager and check whether the sensor is listed. If it isn't recognized, run all pending Windows Updates (including optional ones), temporarily disable antivirus and firewall, restart the PC, and run the sensor driver setup again from the installer flash drive — it's inside the folder named after your sensor's serial number. Then test again.

The image freezes, cuts off, or looks scrambled

This is almost always Windows putting the USB controller to sleep. Open Device Manager → System devices → Intel(R) Management Engine Interface, right-click Properties → Power Management, and uncheck "Allow the computer to turn off this device to save power". Click OK and try again.

The software won't open or asks for a license

Check that the license dongle is plugged in — it must be connected every time you use the software. If it's in and still not recognized, move it to a different USB port and restart the software.

Installing more than one sensor on the same computer

Each Imagen sensor has its own calibration files. If you use two or more sensors on one PC, run the installer once for each sensor's serial number — the calibration for each unit is installed automatically.

Still stuck? Send us the details

Email us with: your sensor's serial number (printed on the flat side of the USB connector), your Windows version, what the error says, whether this is a first installation, and if possible a photo or screenshot of the error. With that, we can usually solve it in one reply.

Support: info@xrddental.com

Make it last for years

Caring for your Imagen sensor

New hygiene sheath for every patientDisposable covers sized for your sensor. Never reuse them.
Disinfect with 70% isopropyl alcoholWipe the sensor and cable with a moistened cloth before each use. No other liquids.
Hold the connector, not the cableWhen plugging and unplugging, always grip the USB connector itself.
Keep the protective cover onBetween uses, store the sensor in its cover, away from static and drops.
Never autoclave, UV or dry heatThe sensor cannot be heat-sterilized — it would be destroyed instantly.
Don't bend, twist or yank the cableCable damage is the most common cause of sensor failure — and it's avoidable.
No bites, drops or impactsThe CMOS chip inside is precise — and fragile. Human or animal, no biting down on the sensor.
Never open the sensorThere are no serviceable parts inside. If anything seems wrong, write to us first.

We've got your back.

More videos, guides and product material are in our Tech Library — and if anything in this setup gives you trouble, one email is all it takes.

Email support Tech Library